DMIS Execution Error Code List
DMIS Execution Error Code List
Error Code | Meaning |
-145/145 | Failed to send the PTMEAS command |
-130,130, -135, -142,-144, -413 | Machine error while measuring a point |
-131,131,-139,-412 | Failed in getting the touch point |
-125,-126,-127,-128,-129, -1131,-21001,-21070, | User cancelled the execution |
-146,-136,-140, -1130,-147,-320, -20321, | Machine didn’t finish the motion properly |
-132,-134,-138,-21030, ,-21055,-21700,-21811, | Machine error |
-137 | Failed in measure the feature in auto-mode |
-1137,-141, -20710 | Failed in measure the point in auto-mode |
-2002, -588,-251,-271,-254,-255, -7006, -7008, | Cannot find an actual |
-2001, -598,-257,-252,-253,-256,-6001,-6006 | Cannot find a nominal |
-2003,-6035 | Cannot add a nominal |
-10000 | Failed in computing a feature |
-1190 | Machine Error/User cancelled execution |
-20000 | Error in measuring the feature in auto-mode |
-30000 | Error in measuring the feature in auto-mode |
-21000 | Execution path is empty |
-20803 | Error in scanning the circle |
-20802, -20314,-20315,-20316,-21910,-317,-318, | Cannot star the motion/Machine error/Finish motion |
-20801 | Error in executing a laser scanning command |
-20800 | Error in measuring a point using the laser sensor |
-20312,-163 | Cannot star the motion |
-20410, -20411 | Cannot measure a point during path execution |
-20420,-21060 | User cancelled the motion during path execution |
-21900 | Problem in calibrating the probe |
-21920 | Encountered a problem in post-calibration |
-21880 | Failed in collecting touch points |
-211 | User chose cancel the search-parameters-dialog |
-1073,-1074 | Failed to calculate the distance |
-271 | Couldn’t output the form profile of a feature |
-301 | Failed to compute the true position of a feature |
-501,-6009,-6010,-6019,-6020,-6036,-6053,-6054, | Cannot find a variable |
-11000 | A problem in executing a tool/tip change operation |
-6001,-6002,-6004,-6005,-6007 | Cannot find a report item |
-6003,-6008 | Cannot find an SPC item |
-6011 | The user cancelled the data entry dialog |
-6012,-6030,-6037,-6038,-6039,-6040,-6051,-6075, | Couldn’t evaluate a variable |
-6023 | Failed to save a variable |
-6025 | Failed to set the value of an integer/double variable |
-6026 | Empty variable value |
-6027 | Failed to find a variable |
-6050,-6052, -6070,-6072 | Cannot find an opened device/file |
-6070 | Encountered an error in executing STDOUT |
-6073 | Couldn’t open a file/device |
-6074 | Couldn’t write to a device/file |
-6077 | Couldn’t evaluate and indexed label |
-7000 | Invalid/Missing JUMP label |
-7001 | The SPC list is empty |
-7002 | Cannot find the SPC variable |
-7003 | User cancelled the execution during the DMIS pause command |
-7004, -7099 | The SPC variable does not exist |
-7010 | Failed in homing the probe head |
-7011 | Failed to load the sensor list |
-7012 | Failed to load the datums list |
-7013 | Sensor does not exist |
-212 | User cancelled table rotation |
-2111 | User cancelled while executing TEXT/Operator command |
-7018 | User cancelled switching to AUTO mode |
-1200 | Improper 5-axis configuration |
-7136 | User cancelled probe-head indexing |
-7175,-7200 | Cannot find an optimum probe |
-7125 | Nominal type does not match the measured-feature type |
-7252 | User cancelled execution while taking a laser shot in manual mode |
-7253 | Failed to take a laser shot in manual mode |
-7254 | Failed in measuring a point in manual mode |
-7255 | Insufficient points for feature calculations |
-7257 | Cannot construct the feature from laser scanned data |
-210 | User cancelled execution while taking a space point |
-131 | Didn’t get all the required measurements (points/laser shots) |
-7132 | Failed in constructing an MGP alignment |
-7137 | Improper constraints for 3dBest-fit alignment calculations |
-7139 | Error in 3dBest-fit alignment calculations (CAD case) |
-7140 | The 3dBest-fit alignment doesn’t achieve the imposed tolerance |
-7141 | Error in 3dBest-fit alignment calculations (Nominals case) |
-7142 | Error in constructing the alignment using 321 method |
-249 | Error in constructing the alignment using CAD-align method |
-7143 | Error in constructing the alignment using simple-321 method |
-8001 | Cannot rotate the alignment (b-value) |
-8002 | Cannot rotate the alignment (a-value) |
-8012 | Alignment does not exist |
-8016 | Failed to construct a curve by best-fit |
-8017 | Failed to construct a surface by best-fit |
-8018 | Failed to construct a feature by best-fit |
-8020, -8021 | Failed to find a feature (actual or nominal) for construction |
-8022 | Failed in intersecting a curve with a geometric feature |
-8023 | Failed in intersecting a plane with another plane |
-8024, -8025 | Failed in intersecting a circle with a line |
-8026 | Failed in intersecting a circle with a sphere |
-8027, -8028 | Failed in intersecting a plane with a sphere |
-8029 | Failed in intersecting circle with a circle |
-8030, -8031 | Failed in finding the minimum circumscribing slot tangent to a curve |
-8032 | Failed in finding the tangency point(s) between a line a circle |
-8033, -8034 | Failed in constructing a line tangent to a circle |
-7201 | User cancelled indexing the probe-head |
-7202 | Sensor does not exist |
-7203 | Machine is not manual mode |
-7204, -7208 | User cancelled moving the probe head to A=0 and B=0 |
-7206 | User cancelled moving the probe head to A=90 and B=180 |
-7209 | User cancelled Auto-calibration-list execution |
-3500 | Unknown DMIS command |
-4000 | Encountered sever problem during DMIS execution |
-7210 | Cannot start the motion |
-40009, -50009 | Sensor doesn’t exist. |
-40010, -50010 | User cancelled probe-head indexing |
-40006, -50006 | Machine error after indexing the probe head |
-40001, -50001 | Machine error in switching to the auto mode |
-40002, -50002 | Error in changing the position speed |
-40003, -50003 | Error in changing the measure speed |
-40004, -50004 | Error in changing the approach distance |
-40005, -50005 | Error in changing the retract distance |
-40008, -50008 | Error in changing the search distance |
-40007, -50007 | Alignment does not exist |
-40011, -50011 | A problem in setting the work-plane |
-8122 | Cannot find an actual needed for setting the work-plane |
-8123 | Cannot find a nominal needed for setting the work-plane |
-15163 | Cannot start a DCC motion during a tool/tip change operation |
-15151 | Cannot finish a DCC motion during a tool/tip change operation |
-15560 | Failed to compare a feature to CAD |
-15700 | Cannot find a suitable sensor that matches the given vector! |
-15310 | Couldn’t finished buffered moves |
-15124 | Failed in executing an arc move |
-15600 | Failed to take a laser shot |
-15350 | Failed in scanning the circle` |
-15351 | Failed in helical scanning |
-15480 | Failed in scanning |
-15321 | The scanning results in no points |
-15790 | Unsupported scanning |
-15791 | A problem happened during scanning |
-15325 | A problem during data collection |
-15197 | The sensor does not exist |
-15527, -15529 | Couldn’t measure a point |
-40300,-40301,-40310,-40320 | User cancelled laser-sensor related command in manual mode |
-40001 | Failed to finish laser scanning properly |
-40002 | Failed start laser scanning |
-40003 | Failed to pause laser scanning |
-40004 | Failed to resume laser scanning |
-40350 | Failed in getting a laser-scan line |